ZWIN-AQMS550 H2S Analyzer

ZWIN-AQMS550 H2S Analyzer

Air Quality Products 27 0

ZWIN-AQMS550 H2S Analyzer has the characteristics of high measurement accuracy, goodreliability, fast response time, easy operation and wide application range. The analyzer supportsmanual calibration and automatic calibration, the calibration program can be set by the user, andsupports remote control and remote upload. Real-time measurement data and instrument statusparameters can be automatically transmitted and queried, which can facilitate relevant depart-ments to grasp the air quality status in a timely and accurate manner.

ZWIN-AQMS550 H2S Analyzer

 

Overview

This product has the characteristics of high measurement accuracy, good reliability, fast response time, easy operation and wide application range. The analyzer supports manual calibration and automatic calibration, the calibration program can be set by the user, and supports remote control and remote upload. Real-time measurement data and instrument status parameters can be automatically transmitted and queried, which can facilitate relevant departments to grasp the air quality status in a timely and accurate manner.This product has the characteristics of high measurement accuracy, good reliability, fast response time, easy operation and wide application range. The analyzer supports manual calibration and automatic calibration, the calibration program can be set by the user, and supports remote control and remote upload. Real-time measurement data and instrument status parameters can be automatically transmitted and queried, which can facilitate relevant departments to grasp the air quality status in a timely and accurate manner.

Technical principle

Catalytic oxidation method Catalytic oxidation method is the measurement of H2S molecules converted into SO2 molecules under high temperature conditions by a specific catalyst.UV fluorescence is based on molecular emission spectroscopy. SO2 molecules absorb ultraviolet (190nm~230nm) energy to become excited state molecules.Excited molecules are unstable and release excess energy in the form of fluorescent photons when returning to the ground state.

 

Advantages

Measuring by ultraviolet fluorescence method, in line with EPA standards

Measurement range: 0-500ppb to 0-20ppm, user selectable

Microprocessor realizes multi-function operation

Software allows monitoring of test data during operation

Continuous self-test with alarm function

Light intensity attenuation self-test function

Bidirectional RS232 port for remote control

Digital state output instrument working parameters

Adaptive signal filtering technology optimizes response time

With temperature compensation and pressure compensation

Color screen display, touch screen operation

Large-capacity memory, automatically store historical data

 

Specifications

Principle:Catalytic Oxidation Method and Ultraviolet Fluorescence Technology

Zero noise: 1 nmol/mol(ppb)(RMS)

Operating temperature:5℃~40℃

Working humidity:0-95%RH

Lower detection limit: <2nmol/mol(ppb)

Zero drift: <5nmol/mol(ppb)/7 days

Range drift: <1 %FS/(7days)

Response time: <5minto90%

Reproducibility: <1%

Instrument size: 178mmx432mmx620mm

Power supply: (200-240)VAC

 

TagAir QualityEnvironmental MonitoringOnline Analyzer Previous: Next: