ZWIN-AQMS550 H2S Analyzer
Overview
This product has the characteristics of high measurement accuracy, good reliability, fast response time, easy operation and wide application range. The analyzer supports manual calibration and automatic calibration, the calibration program can be set by the user, and supports remote control and remote upload. Real-time measurement data and instrument status parameters can be automatically transmitted and queried, which can facilitate relevant departments to grasp the air quality status in a timely and accurate manner.This product has the characteristics of high measurement accuracy, good reliability, fast response time, easy operation and wide application range. The analyzer supports manual calibration and automatic calibration, the calibration program can be set by the user, and supports remote control and remote upload. Real-time measurement data and instrument status parameters can be automatically transmitted and queried, which can facilitate relevant departments to grasp the air quality status in a timely and accurate manner.

Technical principle
Catalytic oxidation method Catalytic oxidation method is the measurement of H2S molecules converted into SO2 molecules under high temperature conditions by a specific catalyst.UV fluorescence is based on molecular emission spectroscopy. SO2 molecules absorb ultraviolet (190nm~230nm) energy to become excited state molecules.Excited molecules are unstable and release excess energy in the form of fluorescent photons when returning to the ground state.
Advantages
Measuring by ultraviolet fluorescence method, in line with EPA standards
Measurement range: 0-500ppb to 0-20ppm, user selectable
Microprocessor realizes multi-function operation
Software allows monitoring of test data during operation
Continuous self-test with alarm function
Light intensity attenuation self-test function
Bidirectional RS232 port for remote control
Digital state output instrument working parameters
Adaptive signal filtering technology optimizes response time
With temperature compensation and pressure compensation
Color screen display, touch screen operation
Large-capacity memory, automatically store historical data
Specifications
Principle:Catalytic Oxidation Method and Ultraviolet Fluorescence Technology
Zero noise: 1 nmol/mol(ppb)(RMS)
Operating temperature:5℃~40℃
Working humidity:0-95%RH
Lower detection limit: <2nmol/mol(ppb)
Zero drift: <5nmol/mol(ppb)/7 days
Range drift: <1 %FS/(7days)
Response time: <5minto90%
Reproducibility: <1%
Instrument size: 178mmx432mmx620mm
Power supply: (200-240)VAC